Volume 7, Issue 1 (Winter 2018)                   J Occup Health Epidemiol 2018, 7(1): 11-19 | Back to browse issues page


XML Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Shabab M, Rismanchian M, Karimi Zeverdegani S, Rangkooy H A. Feasibility evaluation of trace amount of zinc in urine samples using atomic absorption solidified floating organic drop micro-extraction technique. J Occup Health Epidemiol 2018; 7 (1) :11-19
URL: http://johe.rums.ac.ir/article-1-272-en.html

Related article in
Google Scholar

1- Student Research Committee, Department of Occupational Health Engineering, School of Health, Isfahan University of Medical Sciences, Isfahan, Iran.
2- Department of Occupational Health Engineering, School of Health, Isfahan University of Medical Sciences, Isfahan, Iran
3- Department of Occupational Health Engineering, School of Health, Isfahan University of Medical Sciences, Isfahan, Iran. , s_karimi@hlth.mui.ac.ir
4- Departments of Occupational Health, Faculty of Health, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
Article history
Received: 2017/10/24
Accepted: 2018/01/15
ePublished: 2018/04/16
Abstract:   (7026 Views)
Background: The present study was carried out with the aim of extracting trace amounts of zinc in urine samples with ultrasound-assisted emulsification solidified floating organic drop micro-extraction (USAE-SFODME) method by flame atomic absorption spectrometry (FAAS). The efficiency was investigated using the solvent extraction volume, extraction pH, time sonication and temperature extraction. The present study was conducted aiming to respond on the efficiency of SFODME technique in extracting inorganic analytes in biological samples.
Materials and Methods: This was an experimental research with several steps. After preparation standard solution of zinc, USAE-SFODME technique was used for extracting zinc cation from urine samples. This method involves centrifuge, buffer and ligand adding, sonication, extraction of analyte and finally analysis with FAAS. Excel 2010 software was used in this study in order to plot the graphs.
Results: Extraction of zinc was performed under optimized conditions of 2 ml 1-(2-Pyridylazo)-2-naphthol (PAN), 90 µl 1-dodecanol, pH = 5.5, for 20 minutes at 35 ˚C. Recovery, the regression coefficient, and relative standard deviation (RSD) were obtained as 96.6% and 99.0%, respectively. RSD for tree concentration 0.8 µgml-1 Zn cation (Zn2+) was 3.4%. The limit of detection (LOD) was found to be 0.426 µgml-1.
Conclusions: Using green solvents, downsizing the samples, replacement of toxic reagents use, and lack of needing the preparation of the samples are the most important advantages of this technique. USAE-SFODME has a successful development in determining trace amounts of zinc in urine samples which can be performed in chemical laboratories with rather ordinary equipment.
 
Full-Text [PDF 339 kb]   (1655 Downloads) |   |   Full-Text (HTML)  (1317 Views)  

References
1. Al Bakheet SA, Attafi IM, Maayah ZH, Abd-Allah AR, Asiri YA, Korashy HM. Effect of long-term human exposure to environmental heavy metals on the expression of detoxification and DNA repair genes. Environ Pollut 2013; 181:226-32. [DOI] [PMID]
2. King JC, Shames DM, Lowe NM, Woodhouse LR, Sutherland B, Abrams SA, et al. Effect of acute zinc depletion on zinc homeostasis and plasma zinc kinetics in men. Am J Clin Nutr 2001; 74(1):116-24. [DOI] [PMID]
3. Ho M, Heath AL, Baur LA, Cowell CT, Samman S, Garnett SP. Zinc intake in obese adolescents at risk of type 2 diabetes on low-energy diets. J Nutr Intermed Metab 2014; 1:21. [Sciencedirect] [DOI]
4. Flinn JM, Hunter D, Linkous DH, Lanzirotti A, Smith LN, Brightwell J, et al. Enhanced zinc consumption causes memory deficits and increased brain levels of zinc. Physiol Behav 2005; 83(5):793-803. [DOI] [PMID]
5. Barceloux DG. Molybdenum. J Toxicol Clin Toxicol 1999; 37(2):231-7. [DOI] [PMID]
6. Hammond JW. Metal fume fever in the crushed stone industry. J Ind Hyg Toxicol 1944; 26(4):117-19. [Article] [DOI]
7. Lanska DJ, Remler B. Myelopathy among zinc-smelter workers in Upper Silesia during the late 19th century. Neurology 2014; 82(13):1175-9. [DOI] [PMID]
8. Aminian O, Zeinodin H, Sadeghniiat-Haghighi K, Izadi N. Respiratory Symptoms and Pulmonary Function Tests among galvanized workers exposed to zinc oxide. J Res Health Sci 2015; 15(3):159-62. [PMID]
9. King JC, Shames DM, Woodhouse LR. Zinc homeostasis in humans. J Nutr 2000; 130(5S Suppl):1360S-6S. [DOI] [PMID]
10. Martin CJ, Le XC, Guidotti TL, Yalcin S, Chum E, Audette RJ, et al. Zinc exposure in Chinese foundry workers. Am J Ind Med 1999; 35(6):574-80. [PubMed]
11. Abu-El-Halawa R, Zabin SA. Removal efficiency of Pb, Cd, Cu and Zn from polluted water using dithiocarbamate ligands. J Taibah Univ Sci 2017; 11(1):57-65. [Sciencedirect] [DOI]
12. Pena-Pereira F, Lavilla I, Bendicho C. Headspace single-drop microextraction with in situ stibine generation for the determination of antimony (III) and total antimony by electrothermal-atomic absorption spectrometry. Mikrochim Acta 2009; 164(1-2):77-83. [Article] [DOI]
13. Kokosa JM. Recent trends in using single-drop microextraction and related techniques in green analytical methods. Trends Analyt Chem 2015; 71:194-204. [Sciencedirect] [DOI]
14. Mohammadi SZ, Baghelani YM, Mansori F, Shamspur T, Afzali D. Dispersive liquid-liquid microextraction for the simultaneous separation of trace amounts of zinc and cadmium ions in water samples prior to flame atomic absorption spectrometry determination. Quim Nova 2012; 35(1):198-202. [Article] [DOI]
15. Chang Q, Zhang J, Du X, Ma J, Li J. Ultrasound-assisted emulsification solidified floating organic drop microextraction for the determination of trace amounts of copper in water samples. Front Environ Sci Eng 2010; 4(2):187-95. [Article] [DOI]
16. Chen S, Zhu S, Lu D. Solidified floating organic drop microextraction for speciation of selenium and its distribution in selenium-rich tea leaves and tea infusion by electrothermal vapourisation inductively coupled plasma mass spectrometry. Food Chem 2015; 169:156-61. [DOI] [PMID]
17. Liang P, Liu G, Wang F, Wang W. Ultrasound-assisted surfactant-enhanced emulsification microextraction with solidification of floating organic droplet followed by high performance liquid chromatography for the determination of strobilurin fungicides in fruit juice samples. J Chromatogr B Analyt Technol Biomed Life Sci 2013; 926:62-7. [DOI] [PMID]
18. Ma J, Zhang J, Du X, Lei X, Li J. Solidified floating organic drop microextraction for determination of trace amounts of zinc in water samples by flame atomic absorption spectrometry. Microchim Acta 2010; 168(1-2):153-9. [Article] [DOI]
19. Ma J, Lu W, Chen L. Recent advances in dispersive liquid-liquid microextraction for organic compounds analysis in environmental water: a review. Curr Anal Chem 2012; 8(1):78-90. [Article] [DOI]
20. Zhao G, Song S, Wang C, Wu Q, Wang Z. Determination of triazine herbicides in environmental water samples by high-performance liquid chromatography using graphene-coated magnetic nanoparticles as adsorbent. Anal Chim Acta 2011; 708(1-2):155-9. [DOI] [PMID]
21. Szabó L, Herman K, Mircescu NE, Fălămaş A, Leopold LF, Leopold N, et al. SERS and DFT investigation of 1-(2-pyridylazo)-2-naphthol and its metal complexes with Al (III), Mn (II), Fe (III), Cu (II), Zn (II) and Pb (II). Spectrochim Acta A Mol Biomol Spectrosc 2012; 93:266-73. [DOI] [PMID]
22. Viñas P, Campillo N, Andruch V. Recent achievements in solidified floating organic drop microextraction. Trends Analyt Chem 2015; 68:48-77. [Article] [DOI]

Add your comments about this article : Your username or Email:
CAPTCHA

Send email to the article author


Rights and permissions
This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

2024 CC BY 4.0 | Journal of Occupational Health and Epidemiology